For Harvester96 Models: Mach II, Mach III & Mach IV
|
||
|
ALWAYS WEAR GLOVES, SAFETY GLASSES AND PROTECTIVE CLOTHING
BIOLOGICAL:
1. Wash the filter plate and screen
plate assemblies and all reagent inlet lines with at least
2. Prepare a solution consisting of 10% mix of 5.25% Sodium Hypochlorite (Cholorox Bleach)
3. Select program "Rinse 16" and set the following parameters:
4.Run program "Rinse 16" for two (2) cycles using the bleach solution, and three (3) cycles
5. Using an empty reagent container,run the "Pulse Wash" program to remove all fluids from
6.Drain and rinse the Autotrap with Distilled Water to remove the bleach solution from
the valves. RADIOACTIVE *:
1.Follow the above procedure using a decontamination solution (i.e. count off, decon, etc.)
2. Verify equipment has safe level of background counts. |
|||
|
For Harvester96 Models: Mach II M & Mach III M |
|||
|
ALWAYS WEAR GLOVES, SAFETY GLASSES AND PROTECTIVE CLOTHING BIOLOGICAL:
1. Wash the filter plate and screen plate assemblies and all reagent inlet lines with at least
2. Prepare a solution consisting of 10% mix of 5.25% Sodium Hypochlorite (Cholorox Bleach)
3. Run the Harvester96 for 30 seconds using the bleach solution and 45 seconds with Distilled
4. Using an empty reagent container, hold down the Wash/Aspirate button to remove all fluids
5. Drain and rinse the Autotrap with Distilled Water to remove the bleach solution from the valves. RADIOACTIVE *:
1.Follow the above procedure using a decontamination solution (i.e. count off, decon, etc.)
2.Verify equipment has safe level of background counts. |
|||
|
For All Autotraps |
|||
|
ALWAYS WEAR GLOVES, SAFETY GLASSES AND PROTECTIVE CLOTHING BIOLOGICAL:
1. Fill the Autotrap with a solution consisting of 10% mix of 5.25% Sodium Hypochlorite
2. Drain and rinse with Distilled Water to remove the bleach solution from the valves and the RADIOACTIVE *:
1.Follow the above procedure using a decontamination solution (i.e. count off, decon, etc.)
2. Verify equipment has safe level of background counts. |
|||
| |
|
| |